technical data

combi line 3

measured parameterrangeresolution
pO20 – 800 mmHg (SI-units selectable)0.1 mmHg
pCO25 – 200 mmHg (SI-units selectable) 0.1 mmHg
pH6.000 – 8.000 0.001 pH
total-hemoglobin (tHb)4 – 30 g/dl 0.1 g/dl
barometric pressure100 – 900 mmHg
(SI-units selectable)
1.0 mmHg
Na+20 – 250 mmol/l 1.0 mmol/l
K+0 – 20 mmol/l 0.01 mmol/l
Ca++0 – 5.0 mmol/l 0.01 mmol/l
Cl-20 – 250 mmol/l 1.0 mmol/l
input parameterrangeresolution
patient temperature 13 – 43 °C 0.1 °C
hemoglobin (tHb) 0 – 30 g/dl (if not measured)0.1 g/dl
FIO2 15 – 100 % only relevant for AaDO2
RQ 0.7 – 1.0only relevant for AaDO2
calculated parameterrangeresolution
hydrogen ion conc. (H+)10 – 1000 nmol/l 0.1 nmol/l
actual bicarbonate (HCO3-A) 10 – 50 mmol/l0.1 mmol/l
standard bicarbonate (HCO3-S) 10 – 50 mmol/l 0.1 mmol/l
base exess (BE) -25 – 25 mmol/l0.1 mmol/l
base exess (BE ecf) -25 – 25 mmol/l 0.1 mmol/l
total CO2 (TCO2) 10 – 50 mmol/l0.1 mmol/l
buffer base (BB) 0 – 100 mmol/l0.1 mmol/l
O2 saturation of tHb (O2-sat) 20 – 100 % 0.1 %
O2 content or conc. (O2CT) 0 – 40 % 0.1 %
partial O2-press. at 50% O2-sat. (P50) 10 – 50 mmHg0.01 mmHg
alveolar to arterial oxygen-tension grade (AaDO2) 0 – 800 mmHg 0.1 mmHg
anion gap (A-GAP) 0 – 99 mmol/l0.1 mmol/l
SHUNT 0 – 50 %0.1 %
acid base status relevant diagnosis recorded on printer
hematocrite (Hct) 0 – 100 %0.1 %
data inputdata output
capacitive touchscreen display 7” widescreen color LED-backlight display (TFT 800 x 480)
barcode reader printer fast, low-noise thermoprinter
interfacesnet work (LAN), USB, HL7, GDT
data baseintegrated data bases for patient- and QC-data
calibrationCAL-PACK calibration catridges
automatic calibration interval every 90 min
economy mode intervalevery 240 min
specimen
automatic sample aspiration
„auto“-QC CAL-PACK cartridge with integrated QC-solutions (optional)

specimen container capillary, syringe and others
specimen materialwhole blood, serum, plasma / respiratory gas
dimensions | weight electrical data
height552 mm voltage 115 V resp. 230 V
width260 mmfrequency50/60 Hz
depth282 mmconsumption50 W
weightappr. 13 kgambient temperature12 – 32 °C

All statements without guarantee. Subject to change.
with innovative performance features.

download brochure